Transferrin
Beijing Strong Biotechnologies, Inc
TF
Transferrin Assay Kit
Method: Immunoturbidimetric Liquid reagent, R1: R2=3:1
CLINICAL SIGNIFICANCE
The level of transferrin in serum can be altered in accordance with body,s iron requirements and iron reserves, the transferrin iron would increase in plasma when the iron is decreased, the level has good correlation with the total iron binding capacity in the serum. Transferrin increased during pregnancy, controlled by estrogen and other relevant information in: anemia[1], iron deficiency, inflammation, or malignant disease, liver disease[2、3], malnutrition and protein loss.
PRECISION
Intra assay precision |
||
N=20 |
level 1 |
level 2 |
Mean(g/L) |
2.00 |
3.53 |
SD |
0.016 |
0.024 |
CV(%) |
0.80 |
0.68 |
Inter assay precision |
|||
N=5 |
Batch 1 |
Batch 2 |
Batch 3 |
Mean(IU/ml) |
2.00 |
2.04 |
2.04 |
|
2.02 |
||
(Xmax-Xmin)/ |
(2.04-2.00)/2.02*100=1.9 |
||
Correlation
The correlation of a well-known brand(X) and Gcell(Y) is y = 0.8108x - 0.0065, R2 = 0.9993
Interference
Analytes |
Concentration |
Bilirubin |
400 umol/l |
Hemoglobin |
500 mg/dl |
TG |
500 mg/dl |
Linearity
Linear range [0.25,4.99] ng/ml, the linear correlation coefficient r should be≥0.990, in [2.0,4.99] ng/ml range, the mesured linear deviation should be no more than±10%.