Protein C

INTENDED USE

For in vitro quantitative determination of  PC in human plasma.

CLINICAL SIGNIFICANCE

It is mainly used to assist in the diagnosis of deep venous thrombosis and acute pulmonary thromboembolism caused by congenital or acquired protein C deficiency.

INTERFERENCE

When the concentration of bilirubin was no more than 0.25g/L, there was no significant effect on the specimen determination (The relative deviation does not more than ±10%).

Accuracy

Determine the enterprise reference product, and the relative deviation between the measurement result and the label value is ±10%.

Linearity

In the interval of [10%, 140%], the linear correlation coefficient r0.99.

Repeatability

The measurement is repeated 10 times for normal and abnormal samples, and the coefficient of variation (CV%) of the measured value should not be greater than 10%.

Difference between batches

Three batches of reagents are used to determine normal samples, and the coefficient of variation (CV) of the results should not be more than 15%.   